Simplifying Blood Pressure Measurements in Clinical Settings
DOI:
https://doi.org/10.18203/2394-6040.ijcmph20223525Keywords:
Sphygmomanometers, Cuff size, Upper arm circumference, Adjustment factorAbstract
Background: Upper arm sphygmomanometry is the most commonly used method to measure blood pressure in adults. However, variations in upper arm circumference and use of different cuff-sizes results in different pressure readings. When using the same cuff size, pressure readings will be higher for larger arm circumferences and lower for smaller arm circumferences. The objective of this study was to identify an adjustment factor that will allow pressure readings obtained for any combination of arm circumference and cuff size to be compared.
Methods: To investigate the relationship between arm circumferences, cuff size and pressure readings, experiments were conducted using laboratory simulations and blood pressure measurements on nineteen human subjects. Power analysis identified minimum sample size. Results were analyzed using Chi-square and t-tests. The study was conducted between 2019 and 2021 in Boise, Idaho, USA. The University institutional review board approved the use of human subjects.
Results: Simulations revealed a 99% linear correlation between changes in arm circumference coverage and changes in pressure readings. Human subject tests showed a 1% change in upper arm coverage by the sphygmomanometer cuff corresponded to a 1mmHg change in both systolic and diastolic pressure readings.
Conclusions: The proposed adjustment factor can simplify blood pressure measurements in clinical settings by allowing healthcare providers to use only one sphygmomanometer size. It will also provide the basis for a “reference” against which blood pressure values obtained for any combination of cuff size and arm circumference can be standardized.
References
O’Brien, E. Review: A Century of Confusion: Which bladder for accurate blood pressure measurement?. J Hum Hypertens. 1996;10:565-72.
Fonseca-Reyes S, Fajardo-flores I, Montes-Casillas M, Forsyth-Macquarrie A. Differences and effects of medium and large adult cuffs on blood pressure readings in individuals with muscular arms. Blood Press Monit. 2009;14:166-71.
Bakx C, Oerlemans G, van den Hoogen H, van Weel C, Thien T. The influence of cuff size on blood pressure measurement. J Hum Hpertens. 1997;11:439-45.
Sprafka JM, Strickland D, Gomez-Marin O, Prineas RJ. The effect of cuff size on blood pressure measurement in adults. Epidemiol. 1991:214-17.
Pickering TG, Hall JE, Appel LJ. Recommendations for blood pressure measurement in humans: an AHA scientific statement from the council on high blood pressure research professional and public education subcommittee. J Clin Hypertens. 2005;7:102-9.
Manning DM, Kuchirka C, Kaminski J. Miscuffing: inappropriate blood pressure cuff application. Hypertension. 1983;68(4):763-6.
Jones DW, Appel LJ, Sheps SG. Measuring blood pressure accurately: new and persistent challenges. JAMA. 2003;289(8):1027-30.
Marks LA, Groch A. Optimizing cuff width for noninvasive measurement of blood pressure. Blood Press Monitor. 2000;5(3):153-8.
Stolt M, Sjonell G, Astrom H, Hansson L. Factors affecting the validity of the standard pressure cuff. Clin Physiol. 1993;13(6):611-20.
Russell AE, Wing LM, Smith SA, Aylward PE, McRitchie RJ, Hassam RM, et al. Optimal size of cuff bladder for indirect measurement of arterial pressure in adults. J Hypertens. 1989;7(8):607-13.
Prineas RJ. Measurement of blood pressure in the obese. Ann Epidemiol. 1991;1(4):321-6.
Bovet P, Hungerbuhler P, Quilindo J, Grettve ML, Waeber B, Burnand B. Systematic difference between blood pressure readings caused by cuff type. Hypertension. 1994;24(6):786-92.
Croft PR, Cruickshank JK. Blood pressure measurement in adults: large cuffs for all?. Epidemiol Commun Health. 1990;44(2):170-3.
Mattoo TK. Arm cuff in the measurement of blood pressure. Am J Hypertens. 2002;15(2):675-85.
Pickering TG, Hall JE, Appel LJ. Recommendations for blood pressure measurement in humans and experimental animals. Hypertens. 2005;45:142-61.
Marks LA, Groch A. Optimizing cuff width for noninvasive measurement of blood pressure. Blood Press Monit. 2000;5(3):153-8.
Clark JA, Lieh-Lai MW, Sarnaik A, Mattoo TK. Discrepancies between direct and indirect blood pressure measurements using various recommendations for arm cuff selection. Pediatrics. 2002;110(5):920-3.